Troubleshooting History Dialogs
21 - 30 V Series Operating Instructions
21.7 Troubleshooting
MESSAGE/PROBLEM* REASON SOLUTION
Navigation arrows are not
available
There is no data going in the
direction of the selected arrow.
Check the patient’s admit time or
date.
Already at the top or bottom or far
left or far right of a list.
Quick Trends list is empty The list does not populate until an
NIBP measurement is taken.
Take an NIBP measurement.
No Further Events There are no earlier or later event
from where the cursor is currently
positioned.
Select a button to move forward or
backward.
Cannot delete event Event is archived. Uncheck the archived checkbox.
Event cannot be deleted. Certain events cannot be deleted.
Too many archived events! Attempted to archive more than
100 events.
Only able to archive100 events at
one time. Uncheck some events.
Too many tagged events! Attempted to tag more than 100
events.
Only able to tag 100 events at one
time. Uncheck some events.
Can not set the Archive status
for this event!
To prevent deleting, the selected
event cannot be unarchived.
No solution required.
* Messages are shown in all bold text.

Table of Contents

Other manuals for Mindray V21

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray V21 and is the answer not in the manual?

Mindray V21 Specifications

General IconGeneral
BrandMindray
ModelV21
CategoryMedical Equipment
LanguageEnglish

Related product manuals