DIRECTION 5854713-100, REVISION 2
VENUE FIT™ SERVICE MANUAL
5-8 Chapter 5 - Components and Function (Theory)
PRELIMINARY
Section 5-5
Compact Front End Board (C-FEB)
5-5-1 General Information
The main purpose of the Venue Fit™ Front End Board (C-FEB) is to transmit/receive all the channels
data signals to/from the 128 TRX channels on the board. In addition, it contains all the control and status
lines required for HV MUX setup, probe selection, reading probes status, probes information and the
power supplies for transmission and operation of the scanner.
The Venue Fit™ Front End Board is designed to support the cSound SW beam-forming architecture for
128 channels.
5-5-2 C-FEB Interfaces
The External Interfaces from the Front End are:
The Main DC IN power connector that connects the AC/DC Adapter 24 V supply to the scanner.
Optional Power Docking connector for future Power Supply from a smart Cart
The C-FEB is connected to the M-BEB via 2 connectors:
High Speed Data and Control Connector that consist of the following:
PCI-e channel (1 Lane Gen2) for Setup and communication between the BEP and the DSP
located on the C-FEB via PCI Bridge
PCI-e channel (8 lanes Gen2) for Acquired Data Transfer between FE and BE
SMBUS Channel between the BMC and the PMC to transfer BE status information or
communication messages from the scanner application.
Control and Status Signals between the BE BMC and the PMC
Power Connector that consist of the following Supply voltages:
12v Main Supply voltage to the M-BEB and to the Display connected to it.
The C-FEB is connected to the M-PSB via 2 connectors and consist of the following Interfaces:
28 Transmit and Receive System channels
Dedicated I2C channel for PSB Setup and Status Reading
Discrete Control Signals
I2C channel for Probe ID Reading and PSB Debug Information.
LVDS channels for Matrix Probe Setup and Status Reading (Reserved, but not used in M-PSB)
PSB DC Supply
Probe Low voltage and High Voltage Supply
Internal and External HV Mux Control Interface
The C-FEB contains fan controllers for both inlet fans.