DIRECTION 5854713-100, REVISION 2
VENUE FIT™ SERVICE MANUAL
Chapter 5 - Components and Function (Theory) 5-17
PRELIMINARY
The Charger resumes normal charging when the battery temperature is at least below 43° C.
5-9-3-3 Battery Power Specifications
Scanning time, using 3Sc-RS probe with "Cardiac" factory preset in continuous 2D mode is MIN: 60
minutes (with a new fully charged battery).
Scan time in 2D after 48 hrs of standby mode from a new, fully charged battery is | MIN: 30 min.
The system shall be able to remain in standby mode for at least 36 hours, using a fully charged, new
battery.