DIRECTION 5854713-100, REVISION 2
VENUE FIT™ SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-29
PRELIMINARY
Section 7-6
Using a Loaner System during Faulty System Repair
7-6-1 Purpose of this Section
If your system cannot be repaired via the troubleshooting procedures and a GE Repair Depot is
available in your area, contact your local GE service team for requesting a loaner system. When the
loaner system arrives, transfer the data between the systems according to the flow described below.
NOTE: In case of failure to run the below procedure, contact your local GE service representative for further
support.
NOTE: This utility will only work properly if there are at least 150 files on partition E:\
7-6-2 Transferring data to the Loaner System
1.) Open the package and verify the contents:
Software disk-on-key
Network cable
Loaner system
Figure 7-23 Loaner system kit
2.) On the faulty system, perform the following steps:
1.) Shut down the system
2.) Connect the software Disk-On-Key to any of the system’s USB ports.
3.) Turn on the system.
4.) Select Run Loaner Wizard