DIRECTION 5854713-100, REVISION 2
VENUE FIT™ SERVICE MANUAL
Chapter 5 - Components and Function (Theory) 5-19
PRELIMINARY
Section 5-11
Peripherals
5-11-1 Internal Peripheral
5-11-1-1 Wi-Fi Adapter
The Wi-Fi adapter allows the user to connect the system to a wireless network.
The adapter is connected to the back side of the system via a USB port.
5-11-2 External Peripherals
5-11-2-1 Barcode reader
The barcode reader is an option. It allows the user scanning patient ID from a printed barcode label and
scanning needle type barcode.
The barcode reader is connected to the Venue Fit™ system via a USB port located on the rear.
5-11-2-2 Black & White Digital Graphic Printer
The B/W Printer is available as an option. The connection to the system is wireless via a dedicated
adapter made by Sony.
5-11-2-3 ECG
The ECG module is responsible for the acquisition of the ECG Analog signal inputs by the three leads
connected to the patient body.