DIRECTION 5854713-100, REVISION 2
VENUE FIT™ SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-7
PRELIMINARY
Section 7-4
RSVP Service Desktop
7-4-1 Disruptive mode
Disruptive mode is a way to control interruptions to operation of the system. Disruptive mode is required
whenever service performs a function that may disrupt a normal scan.
Disruptive mode is required to run diagnostics, clean presets, and reset the patient database, and turn
on Virtual Console Observation (VCO).
When Disruptive mode is On, all service functionality on the Service desktop is allowed but user
operation of the system may be limited.
When Disruptive mode is Off, some service functionality on the Service desktop is not available
and user operation of the system is normal.
The ability to enable Disruptive mode depends on the logged in user access level:
Local user - a user locally logged into the machine will not be able to set the Disruptive
mode. Local user can only allow a Disruptive mode request from a remote user through
the Service desktop. The local user must have Authorize Remote Service Access to allow
Disruptive mode. If the local user does not have this right, the remote user’s request will
be automatically denied.
Remote user - a user remotely accessing the system will not be able to automatically
switch on the Disruptive mode. The local user has to grant remote access.
The local user will be notified through a dialog box and asked to allow Disruptive mode.