System Setup
3-24 System view Versana Balance – Basic Service Manual
5808768-100 English Rev.9
3-6-6 Connections on the I/O Rear Panel
NOTE: Accessory equipment connected to the analog and digital
interfaces must be certified according to the respective IEC
standards (e.g. IEC60950 for data processing equipment and
IEC60601-1 for medical equipment). Furthermore, all complete
configurations shall comply with the valid version of the system
standard IEC60601-1. Everybody who connects additional
equipment to the signal input part or signal output part of
Versana Balance, configures a medical system, and is therefore
responsible that the Ultrasound system complies with the
requirements of the valid version of IEC60601-1. If in doubt,
consult the technical service department or your local
representative for GE.
3-6-6-1 Connect Ethernet
Connect the network cable to the Ethernet connector on the I/O
Rear Panel.
The connector is located on the rear side of Versana Balance.
3-6-6-2 Connect USB Flash Drive
NOTE: USB Flash Drive approved for Versana Balance are verified for
EMC performance according to EN55011 class B. The use of
any other USB Flash Drive will compromise this verification, and
may cause interference on Versana Balance itself, or on other
electronic devices.
For approved models, please refer to Chapter 9.
Insert the USB Flash Drive in one of the USB ports on the
Versana Balance.
3-6-7 Connecting probes
Please refer to Basic User Manual on how to connect/
disconnect a probe.
3-6-8 Powering the system
Please refer to Basic User Manual on how to power the system.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Versana Balance and is the answer not in the manual?

GE Versana Balance Specifications

General IconGeneral
CategoryUltrasound System
Doppler ModesColor Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler
Power Requirements100-240 V AC, 50/60 Hz
DICOM CompatibilityYes
Wireless ConnectivityOptional
DisplayLED
Transducersconvex, linear, phased array
Image StorageUSB
ConnectivityUSB, HDMI, Ethernet, DICOM
ApplicationsObstetrics, Gynecology, Cardiology, Vascular, Urology
Imaging Modes2D, M-Mode, 3D/4D
Operating Temperature10°C to 40°C
Storage Temperature-20°C to +60°C
Humidity Range20% to 80% (non-condensing)
Display Size15.6 inches
Transducer Ports3 active transducer ports

Related product manuals