General Procedures and Functional Checks
4-54 System view Versana Balance – Basic Service Manual
5808768-100 English Rev.9
4-4-6 Performance Tests
4-4-6-1 Image Quality Check (IQC) preset for service
Image Quality Check (IQC) is intended to facilitate Image
Quality checks during Quality Assurance Evaluation. Quality
Assurance tests are used to determine whether a scanner is
providing the same level of performance year after year.
By using the same setting year after year, this ensures that the
data collection consistent, independently of who performs the
test.
This preset only includes fundamental settings for 2D mode.
Processing modes like SRI, Harmonics, etc., are turned off.
To do IQC, follow the steps below:
NOTE: The IQCforService is only visible when service dongle is
connected to the system.
1. Activate IQC via Utility--> Admin--> System Admin.
Check the “Enable IQC for Service” box. Select Save, then
select Exit.
Figure 4-58. Check “Enable IQC for Service”

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Versana Balance and is the answer not in the manual?

GE Versana Balance Specifications

General IconGeneral
CategoryUltrasound System
Doppler ModesColor Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler
Power Requirements100-240 V AC, 50/60 Hz
DICOM CompatibilityYes
Wireless ConnectivityOptional
DisplayLED
Transducersconvex, linear, phased array
Image StorageUSB
ConnectivityUSB, HDMI, Ethernet, DICOM
ApplicationsObstetrics, Gynecology, Cardiology, Vascular, Urology
Imaging Modes2D, M-Mode, 3D/4D
Operating Temperature10°C to 40°C
Storage Temperature-20°C to +60°C
Humidity Range20% to 80% (non-condensing)
Display Size15.6 inches
Transducer Ports3 active transducer ports