Page #405 background image
System maintenance
Versana Balance – Basic Service Manual System view 10-15
5808768-100 English Rev.9
10-6-3 Physical inspection
NOTE: These features may not be present on all Ultrasound systems.
Table 10-11: Physical checks
Step Item Description
1. Labeling Verify that all Ultrasound system labeling is present and in readable
condition.
2. Scratches & Dents Inspect the exterior for dents, scratches or cracks.
3. Input Power Refer to:
10-6-2-3 ‘Mains cable inspection’ on page 10-13.
4. Cables &
Connectors
Check all internal cable harnesses and connectors for wear and secure
connector seating. Pay special attention to footswitch assembly and probe
strain or bend reliefs.
5. Shielding &
Covers
Check to ensure that all EMI shielding, internal covers, air flow panels and
screws are in place. Missing covers and hardware could cause EMI/RFI
problems while scanning.
6. Control Panel Inspect keyboard and control panel. Note any damaged or missing items.
7. Control Panel
Lighting
Check for proper operation of all operator panel and Freeze Key light.
8. Monitor Inspect the monitor Display for scratches and bad pixels.
Verify proper operation of Contrast and Brightness controls.
Where applicable, confirm that the monitor arm allows:
• swivelling the screen to the left and to the right
• folding the screen to the locked position
• release and adjustment backwards and forwards
• can be adjusted in the up/down positions.
Note: Monitor Arm movement may vary and is not applicable to all
Ultrasound systems.
9. External I/O Check all connectors for damage.
10. Power and
System Status
Indicators
Check for proper operation of all Power and System Status Indicators.
11. Battery Where applicable, check that the battery is not damaged, does not leak,
does not emit an odor, and is not deformed or discolored. Observe all
warnings and cautions for battery handling, recharging, storing, and/or
disposal,

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Versana Balance and is the answer not in the manual?

GE Versana Balance Specifications

General IconGeneral
CategoryUltrasound System
Doppler ModesColor Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler
Power Requirements100-240 V AC, 50/60 Hz
DICOM CompatibilityYes
Wireless ConnectivityOptional
DisplayLED
Transducersconvex, linear, phased array
Image StorageUSB
ConnectivityUSB, HDMI, Ethernet, DICOM
ApplicationsObstetrics, Gynecology, Cardiology, Vascular, Urology
Imaging Modes2D, M-Mode, 3D/4D
Operating Temperature10°C to 40°C
Storage Temperature-20°C to +60°C
Humidity Range20% to 80% (non-condensing)
Display Size15.6 inches
Transducer Ports3 active transducer ports