EMC, EMI and ESD
Versana Balance – Basic Service Manual System view 1-25
5808768-100 English Rev.9
1-10 EMC, EMI and ESD
1-10-1 Electromagnetic Compatibility (EMC)
Electromagnetic compatibility describes a level of performance
of a device within its electromagnetic environment. This
environment consists of the device itself and its surroundings
including other equipment, power sources and persons with
which the device must interface. Inadequate compatibility results
when a susceptible device fails to perform as intended due
interference from its environment or when the device produces
unacceptable levels of emission to its environment. This
interference is often referred to as radio–frequency or
electromagnetic interference (RFI/EMI) and can be radiated
through space or conducted over interconnecting power of
signal cables. In addition to electromagnetic energy, EMC also
includes possible effects from electrical fields, magnetic fields,
electrostatic discharge and disturbances in the electrical power
supply.
1-10-2 CE Compliance
Versana Balance conforms to all applicable conducted and
radiated emission limits and to immunity from electrostatic
discharge, radiated and conducted RF fields, magnetic fields
and power line transient requirements.
For applicable standards, refer to the Safety Chapter of the
Ultrasound system User’s Manual.
NOTE: For CE Compliance, it is critical that all covers, screws,
shielding, gaskets, mesh, clamps, are in good condition,
installed tightly without skew or stress. Proper installation
following all comments noted in this service manual is required
in order to achieve full EMC performance.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Versana Balance and is the answer not in the manual?

GE Versana Balance Specifications

General IconGeneral
CategoryUltrasound System
Doppler ModesColor Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler
Power Requirements100-240 V AC, 50/60 Hz
DICOM CompatibilityYes
Wireless ConnectivityOptional
DisplayLED
Transducersconvex, linear, phased array
Image StorageUSB
ConnectivityUSB, HDMI, Ethernet, DICOM
ApplicationsObstetrics, Gynecology, Cardiology, Vascular, Urology
Imaging Modes2D, M-Mode, 3D/4D
Operating Temperature10°C to 40°C
Storage Temperature-20°C to +60°C
Humidity Range20% to 80% (non-condensing)
Display Size15.6 inches
Transducer Ports3 active transducer ports

Related product manuals