General Procedures and Functional Checks
4-38 System view Versana Balance – Basic Service Manual
5808768-100 English Rev.9
4-3-2 Disk Encryption
NOTE: For R1.0.4 and higher, Encryption is set as “ON” for Local
patient data drive after software installation. User can not turn off
the Disk Encryption. But user can turn on and off the Disk
Encryption for Removable Media.
Figure 4-30. Encryption ON for R1.0.4 and higher
You could change password as you wish.
Select Change Password and you will be prompted to enter
old password/passphrase which you previously set up in EZ
Configuration Wizard, refer to step 3 of
3-9-1 ‘EZ
configuration Wizard’ on page 3-68.
Figure 4-31. Enter Current Password

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Versana Balance and is the answer not in the manual?

GE Versana Balance Specifications

General IconGeneral
CategoryUltrasound System
Doppler ModesColor Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler
Power Requirements100-240 V AC, 50/60 Hz
DICOM CompatibilityYes
Wireless ConnectivityOptional
DisplayLED
Transducersconvex, linear, phased array
Image StorageUSB
ConnectivityUSB, HDMI, Ethernet, DICOM
ApplicationsObstetrics, Gynecology, Cardiology, Vascular, Urology
Imaging Modes2D, M-Mode, 3D/4D
Operating Temperature10°C to 40°C
Storage Temperature-20°C to +60°C
Humidity Range20% to 80% (non-condensing)
Display Size15.6 inches