Electrical Safety Tests Log
Versana Balance – Basic Service Manual System view 10-35
5808768-100 English Rev.9
10-10 Electrical Safety Tests Log
Table 10-19: Electrical safety tests log
Electrical test performed Max value
allowed
Value
measured
OK? Comments
Outlet (correct ground and wiring
config.)
System ground continuity
Chassis source leakage current -
probe
Chassis source leakage current -
wheel
Chassis source leakage current -
monitor
Patient lead source leakage (lead
to ground)
Patient lead source leakage (lead
to lead)
Patient lead source leakage
(isolation)
Peripheral 1 leakage current
Peripheral 1 ground continuity
Peripheral 2 leakage current
Peripheral 2 ground continuity
Peripheral 3 leakage current
Peripheral 3 ground continuity

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Versana Balance and is the answer not in the manual?

GE Versana Balance Specifications

General IconGeneral
CategoryUltrasound System
Doppler ModesColor Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler
Power Requirements100-240 V AC, 50/60 Hz
DICOM CompatibilityYes
Wireless ConnectivityOptional
DisplayLED
Transducersconvex, linear, phased array
Image StorageUSB
ConnectivityUSB, HDMI, Ethernet, DICOM
ApplicationsObstetrics, Gynecology, Cardiology, Vascular, Urology
Imaging Modes2D, M-Mode, 3D/4D
Operating Temperature10°C to 40°C
Storage Temperature-20°C to +60°C
Humidity Range20% to 80% (non-condensing)
Display Size15.6 inches
Transducer Ports3 active transducer ports

Related product manuals