Replacement Procedures
8-4 System view Versana Balance – Basic Service Manual
5808768-100 English Rev.9
8-2-1 Warnings(continued)
NOTE: Use an ESD compatible work space or the ESD-kit during parts
replacement.
8-2-2 Returning/shipping probes and repair parts
Equipment being returned must be clean and free of blood and
other infectious substances.
GE policy states that body fluids must be properly removed from
any part or equipment prior to shipment. GE employees, as well
as customers, are responsible for ensuring that parts/equipment
have been properly decontaminated prior to shipment. Under no
circumstance should a part or equipment with visible body fluids
be taken or shipped from a clinic or site (for example, body coils
or an ultrasound probe). The purpose of the regulation is to
protect employees in the transportation industry, as well as the
people who will receive or open this package.
NOTE: The US Department of Transportation (DOT) has ruled that
“items that were saturated and/or dripping with human blood
that are now caked with dried blood; or which were used or
intended for use in patient care” are “regulated medical waste”
for transportation purposes and must be transported as a
hazardous material.
NOTE: Please remove the battery before system shipment to avoid any
potential danger.
WARNING
The waste of electrical and electronic equipment must not be
disposed as unsorted municipal waste and must be collected
separately.
Please contact the manufacturer or other authorized disposal
company to decommission your equipment.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Versana Balance and is the answer not in the manual?

GE Versana Balance Specifications

General IconGeneral
CategoryUltrasound System
Doppler ModesColor Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler
Power Requirements100-240 V AC, 50/60 Hz
DICOM CompatibilityYes
Wireless ConnectivityOptional
DisplayLED
Transducersconvex, linear, phased array
Image StorageUSB
ConnectivityUSB, HDMI, Ethernet, DICOM
ApplicationsObstetrics, Gynecology, Cardiology, Vascular, Urology
Imaging Modes2D, M-Mode, 3D/4D
Operating Temperature10°C to 40°C
Storage Temperature-20°C to +60°C
Humidity Range20% to 80% (non-condensing)
Display Size15.6 inches
Transducer Ports3 active transducer ports

Related product manuals