System Setup
3-88 System view Versana Balance – Basic Service Manual
5808768-100 English Rev.9
3-9-5 How to get the Versana Balance to recognize another Device on
the Network
Table 3-30: Utility ->Connectivity ->Device screen
1Select Add. The system creates a device called “NewDevice”.
2 Change the name to one of your choosing.
3 Enter the IP address of the device.
4
Save your changes and then press Ping. A “Smiley Face” indicates successful
communication between your Versana Balance and the device. A “Frown” indicates
failed communication. Check the following:
• Is the device running?
• Is it connected to the network?
• Did you enter the right IP address?

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Versana Balance and is the answer not in the manual?

GE Versana Balance Specifications

General IconGeneral
CategoryUltrasound System
Doppler ModesColor Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler
Power Requirements100-240 V AC, 50/60 Hz
DICOM CompatibilityYes
Wireless ConnectivityOptional
DisplayLED
Transducersconvex, linear, phased array
Image StorageUSB
ConnectivityUSB, HDMI, Ethernet, DICOM
ApplicationsObstetrics, Gynecology, Cardiology, Vascular, Urology
Imaging Modes2D, M-Mode, 3D/4D
Operating Temperature10°C to 40°C
Storage Temperature-20°C to +60°C
Humidity Range20% to 80% (non-condensing)
Display Size15.6 inches
Transducer Ports3 active transducer ports

Related product manuals