Disk Encryption/Decryption
Versana Balance – Basic Service Manual System view 4-49
5808768-100 English Rev.9
4-3-6 Change Recovery Key(continued)
3. Press OK.
Figure 4-50. Change Recovery Key 3
Figure 4-51. Change Recovery Key 4
4-3-7 Unlock with user entered key & recovery key
When the patient data is in locked status, you can unlock it with
user entered key and recovery key.
You can insert the USB with recovery key, or you can input the
password or recovery key manually to unlock the system.
NOTE: If you press Cancel, the system will stay in locked status and
some of the functions will be not available.
Figure 4-52. Unlock Dialog

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Versana Balance and is the answer not in the manual?

GE Versana Balance Specifications

General IconGeneral
CategoryUltrasound System
Doppler ModesColor Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler
Power Requirements100-240 V AC, 50/60 Hz
DICOM CompatibilityYes
Wireless ConnectivityOptional
DisplayLED
Transducersconvex, linear, phased array
Image StorageUSB
ConnectivityUSB, HDMI, Ethernet, DICOM
ApplicationsObstetrics, Gynecology, Cardiology, Vascular, Urology
Imaging Modes2D, M-Mode, 3D/4D
Operating Temperature10°C to 40°C
Storage Temperature-20°C to +60°C
Humidity Range20% to 80% (non-condensing)
Display Size15.6 inches
Transducer Ports3 active transducer ports