General Procedures and Functional Checks
4-20 System view Versana Balance – Basic Service Manual
5808768-100 English Rev.9
4-2-10 Exit to Windows Desktop from the Versana Balance
application software
If you want to exit to the Windows Desktop when the application
software is running, please refer to Propretary Service Manual.
4-2-11 Removable media
Refer to the latest revision of the User Manual to perform the
following tasks:
• Using Removable Media
• Labeling Removable Media
Formatting Removable Media
• Verifying Removable Media
4-2-12 Backup and Restore Database, Preset Configurations and
Images
NOTE: Always save presets before any software reload. This ensures
the presets loaded after the software reload are as up–to–date
as possible.
All user presets except changes to Summary, Anatomy, and
Biometry pages, can be saved on an DVD-R disk (or USB
memory device) for reloading on the system.
NOTE: Presets should NOT be saved on the same USB memory device
(or DVD-R disk) as images. The Archive Menu lists the images
but does NOT list the presets stored on a USB memory device
(or DVD-R disk).
NOTE: Restore or Import Patient Archive/User Defined Configuration/
Service/Database are not allowed between R1.0.x and R1.2.x,
as these data are incompatible.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Versana Balance and is the answer not in the manual?

GE Versana Balance Specifications

General IconGeneral
CategoryUltrasound System
Doppler ModesColor Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler
Power Requirements100-240 V AC, 50/60 Hz
DICOM CompatibilityYes
Wireless ConnectivityOptional
DisplayLED
Transducersconvex, linear, phased array
Image StorageUSB
ConnectivityUSB, HDMI, Ethernet, DICOM
ApplicationsObstetrics, Gynecology, Cardiology, Vascular, Urology
Imaging Modes2D, M-Mode, 3D/4D
Operating Temperature10°C to 40°C
Storage Temperature-20°C to +60°C
Humidity Range20% to 80% (non-condensing)
Display Size15.6 inches
Transducer Ports3 active transducer ports

Related product manuals