Service Adjustments
6-6 System view Versana Balance – Basic Service Manual
5808768-100 English Rev.9
6-3-3 Brightness and Contrast
1. Press Cursor key on the control panel. Move the trackball to
position the cursor over the adjustment icon, then press Set
to display the setting menu.
Figure 6-5. Setting Menu Icon
2. Follow the procedure below to adjust the Color
Temperature, Contrast, Brightness and LED Light:
To adjust the color temperature: press Standard,
Brown, or Blue to choose the color temperature. The
chosen color temperature displays in the Color Temp
area.
Figure 6-6. Color Temperature
To adjust the Contrast: Press Set to choose the
adjustment button, hold down the Set key and move the
trackball to move the adjustment button. Move the
adjustment button to the left to decrease the contrast or
move the adjustment button to the right to increase the
contrast.
Figure 6-7. Contrast

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Versana Balance and is the answer not in the manual?

GE Versana Balance Specifications

General IconGeneral
CategoryUltrasound System
Doppler ModesColor Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler
Power Requirements100-240 V AC, 50/60 Hz
DICOM CompatibilityYes
Wireless ConnectivityOptional
DisplayLED
Transducersconvex, linear, phased array
Image StorageUSB
ConnectivityUSB, HDMI, Ethernet, DICOM
ApplicationsObstetrics, Gynecology, Cardiology, Vascular, Urology
Imaging Modes2D, M-Mode, 3D/4D
Operating Temperature10°C to 40°C
Storage Temperature-20°C to +60°C
Humidity Range20% to 80% (non-condensing)
Display Size15.6 inches
Transducer Ports3 active transducer ports