System maintenance
Versana Balance – Basic Service Manual System view 10-17
5808768-100 English Rev.9
10-6-5 Probe maintenance
10-6-5-1 Probe related checks
10-6-5-2 Basic probe care
The Ultrasound system user manuals and various probe
handling cards provide a complete description of probe care,
maintenance, cleaning and disinfection. Ensure that you are
completely familiar with the proper care of GE probes.
Ultrasound probes can be easily damaged by improper
handling. See the User Manual and probe care cards for more
details. Failure to follow these precautions can result in serious
injury and equipment damage. Failure to properly handle or
maintain a probe may also void its warranty.
Any evidence of wear indicates the probe cannot be used.
Do a visual check of the probe pins and Ultrasound system
sockets before plugging in a probe.
The Interoperative probes often have special considerations and
individual probe user manuals. For Interoperative probes also
refer to their separate user manuals.
Table 10-12: System preliminary checks
Step Item Description
1. Probe Holder Clean probe holders. (they may need to be soaked to remove excess gel).
2. Probes Thoroughly check the Ultrasound system probe connectors and remove
dust from inside the connector sockets if necessary. Visually check for
bent, damaged or missing pins.
3. Probes Verify that the Ultrasound system properly recognizes all probes.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Versana Balance and is the answer not in the manual?

GE Versana Balance Specifications

General IconGeneral
CategoryUltrasound System
Doppler ModesColor Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler
Power Requirements100-240 V AC, 50/60 Hz
DICOM CompatibilityYes
Wireless ConnectivityOptional
DisplayLED
Transducersconvex, linear, phased array
Image StorageUSB
ConnectivityUSB, HDMI, Ethernet, DICOM
ApplicationsObstetrics, Gynecology, Cardiology, Vascular, Urology
Imaging Modes2D, M-Mode, 3D/4D
Operating Temperature10°C to 40°C
Storage Temperature-20°C to +60°C
Humidity Range20% to 80% (non-condensing)
Display Size15.6 inches
Transducer Ports3 active transducer ports

Related product manuals