Diagnostics/Troubleshooting
7-4 System view Versana Balance – Basic Service Manual
5808768-100 English Rev.9
7-2-3 Collect a Trouble Image with Logs
If the system should malfunction, press the Alt+D keys
simultaneously. This will collect a screen capture of the image
monitor, system presets and several log files in a date and time
stamped “.zip” file.
NOTE: This function may also be used to make a Print Screen.
This Alt+D function is available at all times.
When Alt+D is pressed, a menu box appears that allows for:
a place to enter a description of the issue
a checkbox to indicate a System lockup
a choice to Export a pre-formated CD-R/DVD-R or save to
the Export directory D: drive (for remote viewing through
FFA)
Figure 7-1. Alt+D Dialog Box

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Versana Balance and is the answer not in the manual?

GE Versana Balance Specifications

General IconGeneral
CategoryUltrasound System
Doppler ModesColor Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler
Power Requirements100-240 V AC, 50/60 Hz
DICOM CompatibilityYes
Wireless ConnectivityOptional
DisplayLED
Transducersconvex, linear, phased array
Image StorageUSB
ConnectivityUSB, HDMI, Ethernet, DICOM
ApplicationsObstetrics, Gynecology, Cardiology, Vascular, Urology
Imaging Modes2D, M-Mode, 3D/4D
Operating Temperature10°C to 40°C
Storage Temperature-20°C to +60°C
Humidity Range20% to 80% (non-condensing)
Display Size15.6 inches
Transducer Ports3 active transducer ports

Related product manuals