Facility needs
Versana Balance – Basic Service Manual System view 2-15
5808768-100 English Rev.9
2-3-3 Required facility needs(continued)
NOTE: Please note that image artifacts can occur, if at any time within
the facility, the ground from the main facility's incoming power
source to the Ultrasound unit is only a conduit.
Dedicated single branch power outlet of adequate
amperage meeting all local and national codes which is
located less than 2.5 m (8 ft.) from the unit’s proposed
location
Door opening is at least 76 cm (30 in) wide
Proposed location for unit is at least 0.5m (1.5 ft.) from the
wall for cooling
Power outlet and place for any external peripheral are within
2 m (6.5 ft.) of each other with peripheral within 1 m of the
unit to connect cables.
Power outlets for other medical equipment.
Power outlets for test equipment within 1 m (3.2 ft.) of
Ultrasound system.
Clean and protected space to store probes (in their cases or
on a rack)
Material to safely clean probes (done with a plastic
container, never metal)
For the amperage requirements, see:
2-2-3 ‘Electrical
requirements’ on page 2-6.
2-3-4 Desirable features
Door is at least 92 cm (3 ft.) wide
Circuit breaker for dedicated power outlet is easily
accessible
Sink with hot and cold water
Receptacle for bio–hazardous waste, like used probe
sheaths
• Emergency oxygen supply
Storage for linens and equipment
Nearby waiting room, lavatory, and dressing room
Dual level lighting (bright and dim)
Lockable cabinet ordered by GE for its software and
proprietary manuals

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Versana Balance and is the answer not in the manual?

GE Versana Balance Specifications

General IconGeneral
CategoryUltrasound System
Doppler ModesColor Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler
Power Requirements100-240 V AC, 50/60 Hz
DICOM CompatibilityYes
Wireless ConnectivityOptional
DisplayLED
Transducersconvex, linear, phased array
Image StorageUSB
ConnectivityUSB, HDMI, Ethernet, DICOM
ApplicationsObstetrics, Gynecology, Cardiology, Vascular, Urology
Imaging Modes2D, M-Mode, 3D/4D
Operating Temperature10°C to 40°C
Storage Temperature-20°C to +60°C
Humidity Range20% to 80% (non-condensing)
Display Size15.6 inches
Transducer Ports3 active transducer ports

Related product manuals