General Procedures and Functional Checks
4-10 System view Versana Balance – Basic Service Manual
5808768-100 English Rev.9
4-2-4 Check System Date and Time
A warning message "Please check the system date and time are
correct" appears on the screen when the system is powered on.
This warning message appears for the possible reasons:
The system is not boot up for over 14 days.
The system time has been changed by 24 hours earlier than
the current system time of last boot-up.
The BIOS time is changed by 24 hours earlier than the
current system by resetting BIOS time,replacing BIOS
module or changing BIOS time.
This warning message is to remind the user to check the system
date in case the system date and time is incorrect.
Figure 4-2. Check system date and time message
Move the cursor to OK and press Cursor key on the control
panel to select OK. The system enters.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Versana Balance and is the answer not in the manual?

GE Versana Balance Specifications

General IconGeneral
CategoryUltrasound System
Doppler ModesColor Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler
Power Requirements100-240 V AC, 50/60 Hz
DICOM CompatibilityYes
Wireless ConnectivityOptional
DisplayLED
Transducersconvex, linear, phased array
Image StorageUSB
ConnectivityUSB, HDMI, Ethernet, DICOM
ApplicationsObstetrics, Gynecology, Cardiology, Vascular, Urology
Imaging Modes2D, M-Mode, 3D/4D
Operating Temperature10°C to 40°C
Storage Temperature-20°C to +60°C
Humidity Range20% to 80% (non-condensing)
Display Size15.6 inches
Transducer Ports3 active transducer ports