General Ultrasound system requirements
Versana Balance – Basic Service Manual System view 2-11
5808768-100 English Rev.9
2-2-5 Probes environmental requirements
2-2-5-1 Operation, storage and transport temperatures for probes
Probes should be operated, stored, or transported within the
parameters outlined below.
NOTE: Refer to
Table 2-1 on page 2-3 to determine the needed
settlement time.
2-2-6 Time and manpower requirements
Site preparation takes time. Begin site preparation checks as
soon as possible, if possible, six weeks before delivery, to allow
enough time to make any changes.
CAUTION
Ensure that the probe face temperature does not exceed the
normal operation temperature range.
Table 2-6: Probe Environmental Requirements
Operational Storage Transport
Temperature 10° - 40° C
50° - 104° F
-5° - 50° C
23° - 122° F
-5° - 50° C
23° - 122° F
Humidity 30- 80%
non-condensing
10 - 90%
non-condensing
10 - 90%
non-condensing
Pressure 700 - 1060hPa 700 - 1060hPa 700 - 1060hPa
CAUTION
Check the room temperature before you use the probe.
CAUTION
Ensure that the probe face temperature does not exceed the
normal operation temperature range.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Versana Balance and is the answer not in the manual?

GE Versana Balance Specifications

General IconGeneral
CategoryUltrasound System
Doppler ModesColor Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler
Power Requirements100-240 V AC, 50/60 Hz
DICOM CompatibilityYes
Wireless ConnectivityOptional
DisplayLED
Transducersconvex, linear, phased array
Image StorageUSB
ConnectivityUSB, HDMI, Ethernet, DICOM
ApplicationsObstetrics, Gynecology, Cardiology, Vascular, Urology
Imaging Modes2D, M-Mode, 3D/4D
Operating Temperature10°C to 40°C
Storage Temperature-20°C to +60°C
Humidity Range20% to 80% (non-condensing)
Display Size15.6 inches
Transducer Ports3 active transducer ports

Related product manuals