General procedures
Versana Balance – Basic Service Manual System view 4-33
5808768-100 English Rev.9
4-2-20 Cleaning the air filters
Clean the system's air filters to ensure that a clogged filter does
not cause the system to overheat and reduce system
performance and reliability. It is recommended the filters be
cleaned every two weeks, but the requirements will vary due to
your system use.
4-2-20-1 Manpower
One person, 5 minutes,
4-2-20-2 Tools
• None
4-2-20-3 Preparations
To get access to the air filter for cleaning, you must perform the
following steps:
1. Power down the system.
2. Disconnect the mains power cable from the wall outlet.
Follow these links if you need more information:
See
4-2-3 ‘Power off’ on page 4-8 for more information.
CAUTION
Be sure to lock the wheels before cleaning the air filters to
avoid injury by any unexpected movement of the system.
DO NOT operate the unit without the air filters in place.
Allow the air filters to dry thoroughly before re-installing them
on the unit.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Versana Balance and is the answer not in the manual?

GE Versana Balance Specifications

General IconGeneral
CategoryUltrasound System
Doppler ModesColor Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler
Power Requirements100-240 V AC, 50/60 Hz
DICOM CompatibilityYes
Wireless ConnectivityOptional
DisplayLED
Transducersconvex, linear, phased array
Image StorageUSB
ConnectivityUSB, HDMI, Ethernet, DICOM
ApplicationsObstetrics, Gynecology, Cardiology, Vascular, Urology
Imaging Modes2D, M-Mode, 3D/4D
Operating Temperature10°C to 40°C
Storage Temperature-20°C to +60°C
Humidity Range20% to 80% (non-condensing)
Display Size15.6 inches
Transducer Ports3 active transducer ports

Related product manuals