Diagnostics/Troubleshooting
7-10 System view Versana Balance – Basic Service Manual
5808768-100 English Rev.9
7-4-3 System voltage failure
Figure 7-10. System voltage failure
Cause: Hardware and/or voltage error detected.
What to do
1. Press OK and reboot the system.
2. if the problem persists, shut down the system, turn OFF the
circuit breaker and then reboot the system.
7-4-4 Hardware configuration error
Figure 7-11. Hardware configuration error
Cause: Hardware(s) is not detected as valid system.
System is able to boot up but not able to scan.
What to do
1. Press OK and reboot the system.
2. If the problem persists, shut down the system, turn OFF the
circuit breaker and then reboot the system.
3. if the problem persists, shut down the system and analyze
the log files in Technical Support Mode.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Versana Balance and is the answer not in the manual?

GE Versana Balance Specifications

General IconGeneral
CategoryUltrasound System
Doppler ModesColor Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler
Power Requirements100-240 V AC, 50/60 Hz
DICOM CompatibilityYes
Wireless ConnectivityOptional
DisplayLED
Transducersconvex, linear, phased array
Image StorageUSB
ConnectivityUSB, HDMI, Ethernet, DICOM
ApplicationsObstetrics, Gynecology, Cardiology, Vascular, Urology
Imaging Modes2D, M-Mode, 3D/4D
Operating Temperature10°C to 40°C
Storage Temperature-20°C to +60°C
Humidity Range20% to 80% (non-condensing)
Display Size15.6 inches
Transducer Ports3 active transducer ports

Related product manuals