General Procedures and Functional Checks
4-56 System view Versana Balance – Basic Service Manual
5808768-100 English Rev.9
4-4-6-2 Test Phantoms
The use of test phantoms is only recommended if required by
your facility's (customer’s) QA program.
4-4-7 B Mode Checks
4-4-7-1 Introduction
The B Mode is the system’s default mode.
4-4-7-2 B Mode Screen Example
4-4-7-3 Preparations
Connect one of the probes.
Turn ON the scanner.
The B Mode is displayed (default mode).
4-4-7-4 Adjust the B mode controls
Press B Mode on the Operator Panel to access B mode.
These Image Controls are used to optimize the B Mode picture.
Verify that all the listed controls are working as intended:
• Use Gain and TGC controls to optimize the overall image
together with the Power control.
• Use Depth to adjust the range to be imaged.
• Use Focus to center the focal point(s) around the region of
interest.
• Use Frequency (move to higher frequencies) or Line
Density (move to higher line density) to increase resolution
in image.
• Use Frequency (move to lower frequency) to increase
penetration.
• Use the Rejection control to reduce noise in the image.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Versana Balance and is the answer not in the manual?

GE Versana Balance Specifications

General IconGeneral
CategoryUltrasound System
Doppler ModesColor Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler
Power Requirements100-240 V AC, 50/60 Hz
DICOM CompatibilityYes
Wireless ConnectivityOptional
DisplayLED
Transducersconvex, linear, phased array
Image StorageUSB
ConnectivityUSB, HDMI, Ethernet, DICOM
ApplicationsObstetrics, Gynecology, Cardiology, Vascular, Urology
Imaging Modes2D, M-Mode, 3D/4D
Operating Temperature10°C to 40°C
Storage Temperature-20°C to +60°C
Humidity Range20% to 80% (non-condensing)
Display Size15.6 inches
Transducer Ports3 active transducer ports

Related product manuals