I
-
I
-
Automating Measurements
Automated Measurement Setup and Control
Using learn Strings to
The
IEEE
488
*LRN
(“learn”) command can also be used to set or query a
Save and Recall
complete set of instrument parameters. This can be used as a programming
Instrument States
convenience, eliminating the need for using disk
hles
when saving and
recalling instrument states.
To
obtain the learn string containing the instrument state, use the command
*LRN? as follows:
10 DIM
Statel$
C40001
20 OUTPUT
cORfna;
“*LRN?”
30 ENTER
QRfna
USING
“-K”;Statel$
. . .
160
! Put the learn string back
170 OUTPUT
ORfna;Statel$
Since
*LRN?,
by
IEEE
definition,
only contains the actual instrument state,
exclusive of data traces and calibration arrays, the network analyzer provides
the command
SYST:SET:LRNLong?
This command saves the data traces and calibration arrays if they are
enabled under
(
SAVE
RECALL)
Define Saws or using the MMEM:STOR: STATE
SCPI commands. Using the
*LRN
command to set the instrument settings
takes about the same amount of time as recalling a
hle
from disk using
MMEM:LOAD:STATe.
For more details on learn strings, refer to “Example Programs” in the
Programmer’s Guide.
7-48
深圳市盛腾仪器仪表有限公司 Tel:0755-83589391 Fax:0755-83539691 Website: www.Sengt.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP 8711C and is the answer not in the manual?

HP 8711C Specifications

General IconGeneral
BrandHP
Model8711C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals