Chapter 7: Repair Procedures
Left Rail Internal Repairs
7-32 Service Manual
7.5.5.2 Reinstalling the Harness on the Left Rail
1. On the left rail end of the replacement harness, mark the location on the harness where the cable tie
was located.
2. Remove the cable tie that secures the harness to the carrier.
3. Open the first 4 cable guard latches.
4. Unsnap and remove the end link from the cable carrier.
5. Thread cable through the left rail and out hole on side
6. Pull cable through the rail until the cable mark is in the correct position just inside rail.
7. Reinstall the bushing in the hole on the side of the rail
8. Verify the mark on the cable is still in the correct position and reconnect the cable carrier to the end link
riveted to the left rail. Close the other 4 cable guard latches.
9. Reinstall new cable tie to secure the harness to the carrier.
10. Reinstall the 3-pin connector.
FIGURE 7-32. Cable Tie
Cable Tie
6600-1097-603
Connector
Blue P1
Green P2
Brown P3
FIGURE 7-33. Connector Drawing

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Giraffe OmniBed Carestation CS1 and is the answer not in the manual?

GE Giraffe OmniBed Carestation CS1 Specifications

General IconGeneral
BrandGE
ModelGiraffe OmniBed Carestation CS1
CategoryMedical Equipment
LanguageEnglish

Related product manuals