Chapter 7: Repair Procedures
Radiant Heater Assembly Repair Procedures
Service Manual 7-39
7.6.3 Heater Door Spool Assembly
Refer to “Figure 7-39 Heater Door Cable Adjustment”.
To access the heater door spool assembly first disassemble the canopy/heater from the device as
described in “Removing the Canopy and Heater Assembly.”
1. Remove the housing bracket from the heater assembly.
2. To replace the spool assembly, use a 2.5 mm hex key to remove the button head screw that secures the
cams and door links. Then use an 8 mm wrench to remove the 2 nuts that hold the spool.