B-1
Appendix B: Specifications
Power Requirements
11.5 A @ 100V ~, 50/60 Hz 2 A @ 100V ~, 50/60 Hz (Accessory Outlet)
9.5 A @ 115V ~, 50/60 Hz 2 A @ 115V ~, 50/60 Hz (Accessory Outlet)
5.5 A @ 220V ~, 50/60 Hz 1 A @ 220V ~, 50/60 Hz (Accessory Outlet)
5.5 A @ 230V ~, 50/60 Hz 1 A @ 230V ~, 50/60 Hz (Accessory Outlet)
5.5 A @ 240V ~, 50/60 Hz 1 A @ 240V ~, 50/60 Hz (Accessory Outlet)
Inrush for 1/2 cycle current < 80 A
Electrical
Operation Continuous
Device Type Class 1
Applied Part Type B
Operating Environment
Temperature 20 - 30°C
Humidity
5 - 85% Non-condensing relative humidity
Altitude Up to 10,000 ft (3,048m, 696 hPa)
Air Velocity Up to 0.3 m/s
Water Ingress IPX0
Use Environment Equipment not suitable for use in the presence of a flammable anesthetic
mixture.
Storage Conditions
Temperature -25 to 60°C
Humidity ≤ 85% Non-condensing relative humidity (above 0°C)
Pressure 50 - 106 kPa
User Control Settings
Patient Control Temperature 35 - 37.5°C in 0.1°C increments
Radiant Heat Power 0 - 100% in 5% increments
Air Control Temperature 20 - 39°C in 0.1°C increments
Humidity 30 - 95% RH in 5% increments
Cleaning
Cold Chemical Disinfectant
Sterilization Not Required
Defibrillation Proof
Applied Parts Area Not Defibrillation Proof
Signal Input/Output
RS-232 Serial Data