Chapter 3: Calibration and Maintenance
Scale Calibration
3-6 Service Manual
Note: One 5 kg Certified Test Weight (part number 6600-0209-800) is required to perform the calibration
procedure.
1. Clear all objects from the bed, and ensure only the mattress and clear plate remain on the scale.
2. Ensure the bed is level.
3. In user mode, touch Setup on the Home screen. The Setup screen appears.
4. Touch the data Baby Name field Enter here. The keyboard appears.
5. If this is the first time accessing the application, enter Pass0000 and go to step 6. If not, enter your
password (password + PIN). The password is not case sensitive and may be typed in upper or lower
case.
6. Wait for Password Verified to be displayed.
7. Turn off the Standby switch to turn off the device.
8. Turn on the Standby switch within one minute to turn on the device. The device restarts in service mode
and is open to the Setup screen.
9. In service mode, touch Calibration in the left-navigation area. The display area is populated.
10. Touch Cal Scale.
11. When the screen prompts Remove the Weight and Press Next, verify the weight is removed and press
Next.
A series of instructions appears on-screen.
FIGURE 3-3. Cal Scale

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Giraffe OmniBed Carestation CS1 and is the answer not in the manual?

GE Giraffe OmniBed Carestation CS1 Specifications

General IconGeneral
BrandGE
ModelGiraffe OmniBed Carestation CS1
CategoryMedical Equipment
LanguageEnglish

Related product manuals