Chapter 4: Understanding the Service Interface
Accessing Service Mode
4-2 Service Manual
3. If this is the first time accessing the application, enter Pass0000 and go to step 6. If not enter your
password (Pass + PIN). The password is not case sensitive and may be typed in upper or lower case.
Note: If this is the first time you are accessing the application, you have the option to change your
password as described in “Change Password” on page 4-2.
4. Wait for Password Verified to display.
5. Turn off the Standby switch to turn off the device.
6. Turn on the Standby switch within one minute to turn on the device.
The device restarts in Service Mode and is open to the Setup screen.
4.2.1 Change Password
You have the option to change your password upon initial access to the Service application. You may also
change it at any time.
To change your password:
1. Access service mode.
2. Touch Reset Password.
3. Enter a 4 digit personal identification number (PIN) on the keypad.
4. Reenter the PIN.
5. Touch Yes to set the new password and No to cancel.
Note: The new password (not case sensitive) is Pass + your 4 digit PIN.
4.2.2 Forgotten Passwords
If you forget your password, the only way to get into the service application is through hardware.
To reset a forgotten password:
1. Power off the device.
2. Remove Rear Panel Cover to access the Control Board.
3. Partially slide out the Control Board to expose the Dip Switch, S1.
4. Set Dip Switch S1 position 8 to ON.
5. Power on the device
You are automatically placed in Service mode.
6. Touch Reset Password.
7. Set a new password.
8. Power off the device
9. Set Dip Switch S1 position 8 to OFF.
10. Slide Control Board back to original position.
11. Replace Rear Panel Cover.
12. Power on the device.
The device starts in operator mode.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Giraffe OmniBed Carestation CS1 and is the answer not in the manual?

GE Giraffe OmniBed Carestation CS1 Specifications

General IconGeneral
BrandGE
ModelGiraffe OmniBed Carestation CS1
CategoryMedical Equipment
LanguageEnglish

Related product manuals