EasyManuals Logo
Home>GE>Medical Equipment>Giraffe OmniBed Carestation CS1

GE Giraffe OmniBed Carestation CS1 Service Manual

GE Giraffe OmniBed Carestation CS1
308 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #201 background imageLoading...
Page #201 background image
Chapter 7: Repair Procedures
Display Module Repairs
Service Manual 7-65
7.19 Display Module Repairs
7.19.1 Rear Cover
1. Using a 3 mm hex key, remove the 4 (M4 X 10) button head socket screws, M4 flat washer, and M4 split
washers that secure the back cover to the display module and remove the cover.
7.19.1.1 Carrier Plate Hardware
Location Qty Part Number Description Purpose
A 6 2066776-001 PLASTIC THREAD FORMING
SCREW 40 x 12
FRONT BEZEL MTG
B 4 2066775-001 PLASTIC THREAD FORMING
SCREW 30 x 10
LCD STAND OFF MTG
C 14 2066774-001 THREAD FORMING SCREW
M3 x 8
SBC and ISOLATION SHIELD
MTG
D 2 6600-0713-431 STAR WAHSER M3 ISOLATION SHIELD MTG
FIGURE 7-58. Rear Cover