3: Functions and features Series 2600B System SourceMeter® Instrument
Reference Manual
3-40 2600BS-901-01 Rev. B / May 2013
The programming example below illustrates how to configure a 10-point linear voltage sweep on SMU
A, where each step is triggered by the front-panel TRIG key:
-- Configure a 10-point source voltage sweep.
smua.trigger.source.linearv(10, 100, 10)
smua.trigger.source.action = smua.ENABLE
-- Configure TRIG key press as input trigger for source action.
smua.trigger.source.stimulus = display.trigger.EVENT_ID
-- Command SMU to execute a single 10-point sweep.
smua.trigger.count = 10
smua.trigger.arm.count = 1
-- Turn on the output in preparation for the sweep
smua.source.output = smua.OUTPUT_ON
-- Start the sweep and clear the event detectors.
smua.trigger.initiate()
-- The SMU will wait for the front panel TRIG key press before executing
-- each source action.
-- Wait for the sweep to complete.
waitcomplete()
The following figure graphically illustrates this example. See Sweep operation (on page 3-20) for
more information about sweep operation.
Figure 65: Front panel TRIG key triggering
Using trigger events to start actions on trigger objects
Trigger objects can be configured to respond to events generated by other trigger objects, such as
using a digital I/O trigger to initiate a sweep. To configure a trigger object to monitor for an event,
assign the event ID of the trigger event to the stimulus input. When the specified trigger event occurs,
the trigger object will perform an action. The programming example below illustrates how to generate
a digital I/O line 2 output trigger pulse for each SMU A source complete event:
-- Configure digio line 2 to generate an output trigger pulse each
-- time SMU A generates a source complete event.
digio.trigger[2].stimulus = smua.trigger.SOURCE_COMPLETE_EVENT_ID
The following figure illustrates this example.

Table of Contents

Other manuals for Keithley 2601B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2601B and is the answer not in the manual?

Keithley 2601B Specifications

General IconGeneral
BrandKeithley
Model2601B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals