Series 2600B
System SourceMeter® Instrument Reference Manual Appendix A: Ma
intenance
2600BS-901-01 Rev. B / May 2013 A-5
Using TSB for upgrading the firmware
After downloading the new flash file from the Keithley Instruments website, you can use Test Script
Builder (TSB) to upgrade the firmware of your Series 2600B.
1. If not already running, start Test Script Builder (on the PC desktop, double-click the icon for the
Test Script Builder).
2. On the Instrument Console toolbar, click the Open Instrument icon and then select your
communication interface from the Select Instrument Resource dialog box. See the section on
TSP Programming Fundamentals for details on opening communications.
Figure 143: Open Instrument icon
3. On the Instrument Console toolbar, click the Flash memory icon to display a drop-down menu.
Figure 144: Flash memory icon
4. From the drop-down menu, select Instrument and then click Flash.
5. From the Select A Firmware Data File dialog box, use the browser to select the file name of the
new firmware and click Open to upgrade the firmware.

Table of Contents

Other manuals for Keithley 2601B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2601B and is the answer not in the manual?

Keithley 2601B Specifications

General IconGeneral
BrandKeithley
Model2601B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals