Appendix
B: Calibration Series 2600B System SourceMeter® Instrument
Reference Manual
B-4 2600BS-901-01 Rev. B / May 2013
Verification limits
The verification limits stated in this section have been calculated using only the Series 2600B one-
year accuracy specifications, and they do not include test equipment uncertainty. If a particular
measurement falls outside the allowable range, recalculate new limits based both on the Series
2600B specifications and corresponding test equipment specifications.
Source limit calculations
As an example of how verification limits are calculated, assume you are testing the Model
2601B/2602B/2604B 6 V DC output range using a 5.4 V output value. Using the Model
2601A/2602A/2604B one-year accuracy specification for 5.4 V DC output of ± (0.02% of output + 1.8
mV offset), the calculated output limits are:
Output limits = 5.4 V ± [(5.4 V × 0.02%) + 1.8 mV]
Output limits = 5.4 V ± (0.00108 V + 0.0018 V)
Output limits = 5.4 V ± 0.00288 V
Output limits = 5.39712 V to 5.40288 V
Similarly, assume you are testing the Model 2611B/2612B/2614B/2634B/2635B/2636B 20V DC
output range using an 18 V output value. Using the Model 2611B/2612B/2614B/2634B/2635B/2636B
one-year accuracy specification for 18 V DC output of ± (0.02% of output + 5 mV offset), the
calculated output limits are:
Output limits = 18 V ± [(18 V × 0.02%) + 5 mV]
Output limits = 18 V ± (0.0036 V + 0.005 V)
Output limits = 18 V ± 0.0086 V
Output limits = 17.9914 V to 18.0086 V
Measurement limit calculations
Measurement limits are calculated in the same way as the source limits, except that the limits are
calculated with respect to the measurement of the external reference instrument.
Restoring factory defaults
Before performing the verification procedures, restore the instrument to its factory front panel (bench)
defaults as follows:
1. Press the MENU key.
2. Scroll to the SETUP menu item (by turning the navigation wheel ), and then press the ENTER
key.
3. Scroll to the RECALL menu item, and then press the ENTER key.
4. Scroll to the INTERNAL menu item, and then press the ENTER key.
5. Scroll to the FACTORY menu item.
6. Press the ENTER key to restore defaults.

Table of Contents

Other manuals for Keithley 2601B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2601B and is the answer not in the manual?

Keithley 2601B Specifications

General IconGeneral
BrandKeithley
Model2601B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals