System SourceMeter® Instrument Reference Manual Section 7:
2600BS-901-01 Rev. B / May 2013 7-203
Also see
Measure auto delay (on page 2-83)
reset() (on page 7-160)
smuX.measure.count (on page 7-201)
smuX.measure.delayfactor (on page 7-203)
smuX.source.delay (see "<sm.source.delay" on page 7-220)
smuX.reset() (on page 7-216)
smuX.measure.delayfactor
This attribute stores a multiplier to the delays that are used when smuX.measure.delay is set to
smuX.DELAY_AUTO.
Type TSP-Link accessible Affected by Where saved Default value
SMU reset
Recall setup
Usage
delayFactor = smuX.measure.delayfactor
smuX.measure.delayfactor = delayFactor
The delay factor multiplier
Source-measure unit (SMU) channel (for example,
applies to SMU channel A)
Details
The delay factor is only applied when smuX.measure.delay = smuX.DELAY_AUTO.
This attribute can be set to a value less than 1 (for example, 0.5) to decrease the automatic delay.
This attribute can be set to a value greater than 1 (for example, 1.5 or 2.0) to increase the automatic delay.
Setting this attribute to zero disables delays when smuX.measure.delay = smuX.DELAY_AUTO.
Example
smua.measure.delayfactor = 2.0
Doubles the measure delay for SMU channel A.
Also see
Measure auto delay (on page 2-83)
reset() (on page 7-160)
smuX.measure.delay (on page 7-202)
smuX.reset() (on page 7-216)
smuX.measure.filter.count
This command sets the number of measured readings that are required to yield one filtered measurement.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes
Instrument reset
SMU reset
Saved setup 1