System SourceMeter® Instrument Reference Manual Section 7:
2600BS-901-01 Rev. B / May 2013 7-79
Also see
display.smuX.limit.func (on page 7-77)
display.trigger.clear()
This function clears the front-panel trigger event detector.
Type TSP-Link accessible
Affected by Where saved Default value
Usage
display.trigger.clear()
Details
The trigger event detector remembers if an event has been detected since the last display.trigger.wait()
call. This function clears the trigger event detector and discards the previous history of TRIG key presses.
This attribute also clears the display.trigger.overrun attribute.
Also see
display.trigger.overrun (on page 7-79)
display.trigger.wait() (on page 7-80)
display.trigger.EVENT_ID
This constant is the event ID of the event generated when the front-panel TRIG key is pressed.
Type TSP-Link accessible
Affected by Where saved Default value
Usage
eventID = display.trigger.EVENT_ID
Details
Set the stimulus of any trigger event detector to the value of this constant to have it respond to front-panel trigger
key events.
Also see
None
display.trigger.overrun
This attribute contains the event detector overrun status.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (R) Yes
Instrument reset
Not saved false