EasyManuals Logo
Home>Keithley>Measuring Instruments>2601B

Keithley 2601B Reference Manual

Keithley 2601B
864 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #711 background imageLoading...
Page #711 background image
Series 2600B
System SourceMeter® Instrument Reference Manual Section 7:
TSP command reference
2600BS-901-01 Rev. B / May 2013 7-371
Details
This command reads available data from the remote instrument and returns responses for the specified number
of arguments.
The format string can contain the following specifiers:
%[width]s
Read data until the specified length
%[max width]t
Read data until the specified length or until punctuation is found, whichever comes first
%[max width]n
Read data until a newline or carriage return
%d
Read a number (delimited by punctuation)
A maximum of 10 format specifiers can be used for a maximum of 10 return values.
If formatString is not provided, the command returns a string that contains the data until a new line is
reached. If no data is available, the Series 2600B pauses operation until the requested data is available or until a
timeout error is generated. Use tspnet.timeout to specify the timeout period.
When the Series 2600B reads from a TSP-enabled remote instrument, the Series 2600B removes Test Script
Processor (TSP
®
) prompts and places any errors it receives from the remote instrument into its own error queue.
The Series 2600B prefaces errors from the remote device with "Remote Error," followed by the error number
and error description.
Example
tspnet.write(deviceID, "*idn?\r\n")
print("write/read returns:", tspnet.read(deviceID))
Send the "*idn?\r\n" message to
the instrument connected as
deviceID.
Display the response that is read from
deviceID (based on the *idn?
message).
Also see
tspnet.connect() (on page 7-367)
tspnet.readavailable() (on page 7-371)
tspnet.timeout (on page 7-373)
tspnet.write() (on page 7-376)
tspnet.readavailable()
This function checks to see if data is available from the remote device.
Type TSP-Link accessible
Affected by Where saved Default value
Function
No
Usage
bytesAvailable = tspnet.readavailable(connectionID)
bytesAvailable
The number of bytes available to be read from the connection
connectionID
The connection ID returned from
tspnet.connect()
Details
This command checks to see if any output data is available from the device. No data is read from the instrument.
This allows TSP scripts to continue to run without waiting on a remote command to finish.

Table of Contents

Other manuals for Keithley 2601B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2601B and is the answer not in the manual?

Keithley 2601B Specifications

General IconGeneral
BrandKeithley
Model2601B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals