7: TSP command reference Series 2600B System SourceMeter® Instrument
Reference Manual
7-306 2600BS-901-01 Rev. B / May 2013
status.questionable.calibration.*
This attribute contains the questionable status calibration summary register set.
Type TSP-Link accessible Affected by Where saved Default value
Attribute
- -
- -
- -
- -
.condition (R)
Yes
Not applicable
Not saved
Not applicable
.enable (RW)
Yes
Status reset
Not saved
0
.event (R) Yes Status reset Not saved 0
.ntr (RW)
Yes
Status reset
Not saved
0
.ptr (RW)
Yes
Status reset
Not saved
Models 2601B/2611B/2635B: 2 (All bits set)
Models
2602B/2604B/2612B/2614B/2634B/2636B: 6
(All bits set)
Usage
questionableRegister = status.questionable.calibration.condition
questionableRegister = status.questionable.calibration.enable
questionableRegister = status.questionable.calibration.event
questionableRegister = status.questionable.calibration.ntr
questionableRegister = status.questionable.calibration.ptr
status.questionable.calibration.enable = questionableRegister
status.questionable.calibration.ntr = questionableRegister
status.questionable.calibration.ptr = questionableRegister
questionableRegister
The status of the questionable status calibration summary register; a zero
(0) indicates no bits set (also send 0 to clear all bits); other values indicate
various bit settings
Details
These attributes are used to read or write to the questionable status calibration summary registers. Reading a
status register returns a value. The binary equivalent of the returned value indicates which register bits are set.
The least significant bit of the binary number is bit B0, and the most significant bit is bit B15.
For information about .condition, .enable, .event, .ntr, and .ptr registers, refer to Status register set contents (on
page E-1) and Enable and transition registers (on page E-19). The individual bits of this register are defined in
the following table.
Bit Value Description
B0
Not used
Not applicable.
B1
status.questionable.calibration.SMUA
Set bit indicates that the SMU A calibration
constants stored in nonvolatile memory were
corrupted and could not be loaded when the
instrument powered up.
Bit B1 decimal value: 2
Binary value: 0000 0010
B2
status.questionable.calibration.SMUB
This bit is only available on Models
2602B/2604B/2612B/2614B/2634B/2636B. Set bit
indicates that the SMU B calibration constants
stored in nonvolatile memory were corrupted and
could not be loaded when the instrument powered
up.
Bit B2 decimal value: 4
Binary value: 0000 0100
B3-B15 Not used Not applicable.

Table of Contents

Other manuals for Keithley 2601B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2601B and is the answer not in the manual?

Keithley 2601B Specifications

General IconGeneral
BrandKeithley
Model2601B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals