System SourceMeter® Instrument Reference Manual Section 7:
2600BS-901-01 Rev. B / May 2013 7-43
ConfigPulseVMeasureISweepLog()
This KIPulse factory script
1
function configures a voltage pulse train with a current measurement at each point.
Type TSP-Link accessible
Affected by Where saved Default value
Usage
f, msg = ConfigPulseVMeasureISweepLog(smu, bias, start, stop, limit, ton, toff,
points, buffer, tag, sync_in, sync_out, sync_in_timeout, sync_in_abort)
f, msg = ConfigPulseVMeasureISweepLog(smu, bias, start, stop, limit, ton, toff,
points, buffer, tag, sync_in, sync_out, sync_in_timeout)
f, msg = ConfigPulseVMeasureISweepLog(smu, bias, start, stop, limit, ton, toff,
points, buffer, tag, sync_in, sync_out)
f, msg = ConfigPulseVMeasureISweepLog(smu, bias, start, stop, limit, ton, toff,
points, buffer, tag, sync_in)
f, msg = ConfigPulseVMeasureISweepLog(smu, bias, start, stop, limit, ton, toff,
points, buffer, tag)
A Boolean flag; this flag is
when the pulse was successfully configured,
false when errors were encountered
A string message; if the f flag is false, msg contains an error message; if it is
true,
msg
contains a string indicating successful configuration
System SourceMeter
instrument channel (for example, smua refers to SMU
Pulse sweep start level in volts
Pulse sweep stop level in volts
Current limit (for example, compliance) in amperes
Pulse off time in seconds
Number of pulse-measure cycles
Reading buffer where pulsed measurements will be stored; if this is nil when the
function is called, no measurements will be made when the pulse train is initiated
Numeric identifier to be assigned to the defined pulse train
Defines a digital I/O trigger input line; if programmed, the pulse train will wait for a
trigger input before executing each pulse
Defines a digital I/O trigger output line; if programmed, the pulse train will generate a
trigger output immediately before the start of
Specifies the length of time (in seconds) to wait for input trigger; default value is 10 s
Specifies whether or not to abort pulse if input trigger is not received; if pulse aborts
because of a missed trigger, a timer timeout message is returned;
or
1
The KIPulse factory script provides examples of how to generate pulses and to provide a simple pulsing
interface. Pulses can be generated using the functions listed below.@Please note the following information about
the KIPulse factory script:This factory script only operates on the channels present in the instrument executing the
pulse functions. These functions will not operate correctly if you attempt to access instrument channels over the
TSPLink® interface.The KIPulse factory scripts are ge ...