7: TSP command reference Series 2600B System SourceMeter® Instrument
7-356 2600BS-901-01 Rev. B / May 2013
tsplink.trigger[N].assert()
This function simulates the occurrence of the trigger and generates the corresponding event ID. This function is
not available on the Models 2604B/2614B/2634B.
Type TSP-Link accessible
Affected by Where saved Default value
Usage
tsplink.trigger[N].assert()
The trigger line (1 to 3)
Details
The set pulse width determines how long the trigger is asserted.
Example
tsplink.trigger[2].assert()
Asserts trigger on trigger line 2.
Also see
tsplink.trigger[N].clear() (on page 7-356)
tsplink.trigger[N].mode (on page 7-358)
tsplink.trigger[N].overrun (on page 7-359)
tsplink.trigger[N].pulsewidth (on page 7-360)
tsplink.trigger[N].release() (on page 7-360)
tsplink.trigger[N].stimulus (on page 7-362)
tsplink.trigger[N].wait() (on page 7-364)
tsplink.trigger[N].clear()
This function clears the event detector for a trigger. This function is not available on the Models
2604B/2614B/2634B.
Type TSP-Link accessible
Affected by Where saved Default value
Usage
tsplink.trigger[N].clear()
The trigger line (1 to 3)
Details
The trigger event detector enters the detected state when an event is detected.
tsplink.trigger[N].clear() clears a trigger event detector, discards the history of the trigger line, and
clears the tsplink.trigger[N].overrun attribute.