Series 2600B
System SourceMeter® Instrument Reference Manual Appendix G: Model 24
00 emulation
2600BS-901-01 Rev. B / May 2013 G-17
:SYSTem subsystem
:SYSTem:AZERo:CACHing:NPLCycles?
Always returns 0.
:SYSTem:AZERo:CACHing:REFResh
This command is accepted and ignored. Causes no action or response.
:SYSTem:AZERo:CACHing:RESet
This command is accepted and ignored. Causes no action or response.
:SYSTem:AZERo:CACHing[:STATe] <Bool>
This command is accepted and ignored. The setting is always ON.
:SYSTem:AZERo:CACHing[:STATe]?
Always returns 1.
:SYSTem:AZERo[:STATe] <Bool>
When AZERO state is set to ON, reference and zero measurements will be automatically taken when they are out of
date. If this happens, the time to take a measurement will increase. This can cause sweep timing to be irregular when
compared to a Model 2400 as the Model 2400 takes longer on every measurement.
:SYSTem:POSetup RST|PRESet|SAV0|SAV1|SAV2|SAV3|SAV4
SYSTem:POSetup command is accepted but has no effect, *SAV and *RCL are not supported.
:TRIGger subsystem
:ARM[:SEQuence[1]][:LAYer[1]][:TCONfigure][:ASYNchronous]:ILINe 1|2|3|4
The instrument will not allow the same TLink line to be used for an input trigger and an output trigger simultaneously.
Also, it will not allow the same TLink line to be used as an input trigger for both the arm and trigger layer
simultaneously. The instrument will generate an error when attempting to leave the idle layer if these conditions are
violated. The reset default for the arm layer input line (
ARM:ILINe
) is 1.
:ARM[:SEQuence[1]][:LAYer[1]][:TCONfigure][:ASYNchronous]:OLINe 1|2|3|4
The instrument will not allow the same TLink line to be used for an input trigger and an output trigger simultaneously.
The instrument will generate an error when attempting to leave the idle layer if this condition is violated. The reset
default value for the arm layer output line (ARM:OLINe) is 3.
:TRIGger[:SEQuence[1]][:TCONfigure][:ASYNchronous]:ILINe 1|2|3|4
The instrument will not allow the same TLink line to be used for an input trigger and an output trigger simultaneously
nor will it allow the same TLink line to be used as an input trigger for both the arm and trigger layer simultaneously.
The instrument will generate an error when attempting leave the idle layer if these conditions are violated. The reset
default for the trigger layer input line (
TRIGger:ILINe
) is 2.
:TRIGger[:SEQuence[1]][:TCONfigure][:ASYNchronous]:INPut SOURce|DELay|SENSe|NONE
Only one of these options may be selected at a time.
:TRIGger[:SEQuence[1]][:TCONfigure][:ASYNchronous]:OLINe 1|2|3|4
The instrument will not allow the same TLink line to be used for an input trigger and an output trigger simultaneously.
The instrument will generate an error when attempting to leave the idle layer if this condition is violated. The reset
default for the trigger layer output line (
TRIGger:OLINe
) is 4.

Table of Contents

Other manuals for Keithley 2601B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2601B and is the answer not in the manual?

Keithley 2601B Specifications

General IconGeneral
BrandKeithley
Model2601B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals