Section
7: TSP command reference Series 2600B System SourceMeter® Instrument
Reference Manual
7-222 2600BS-901-01 Rev. B / May 2013
Usage
highC = smuX.source.highc
smuX.source.highc = highC
highC
The state of the high-capacitance mode; set to one of the following values:
0 or smuX.DISABLE: Disables high-capacitance mode
1 or smu
X
.ENABLE: Enables high-capacitance mode
X
Source-measure unit (SMU) channel (for example, smua.source.highc applies
to SMU channel A)
Details
When enabled, the high-capacitance mode has the following effects on the SMU settings:
• smuX.measure.autorangei is set to smuX.AUTORANGE_FOLLOW_LIMIT and cannot be changed
• Current ranges below 1 µA are not accessible
• If smuX.source.limiti is less than 1 µA, it is raised to 1 µA
• If smuX.source.rangei is less than 1 µA, it is raised to 1 µA
• If smuX.source.lowrangei is less than 1 µA, it is raised to 1 µA
• If smuX.measure.lowrangei is less than 1 µA, it is raised to 1 µA
Example
smua.source.highc = smua.ENABLE
Activates high-capacitance mode
for SMU channel A.
Also see
High-capacitance mode (on page 3-65)
smuX.source.levelY
This attribute sets the source level.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW)
Yes
SMU reset
Instrument reset
Recall setup
Saved setup
0
Usage
sourceLevel = smuX.source.levelY
smuX.source.levelY = sourceLevel
sourceLevel
The source value; set to one of the following values:
Voltage: 0 V to ±40 V (Models 2601B/2602B/2604B)
Voltage: 0 V to ±200 V (Models 2611B/2612B/2614B/2634B/2635B/2636B)
Current: 0 A to ±3 A (Models 2601B/2602B/2604B)
Current: 0 A to ±1.5 A (Models 2611B/2612B/2614B/2634B/2635B/2636B)
X
Source-measure unit (SMU) channel (for example, smua.source.levelv applies
to SMU channel A)
Y
SMU source function (v = voltage, i = current)

Table of Contents

Other manuals for Keithley 2601B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2601B and is the answer not in the manual?

Keithley 2601B Specifications

General IconGeneral
BrandKeithley
Model2601B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals