System SourceMeter® Instrument Reference Manual Section 7:
2600BS-901-01 Rev. B / May 2013 7-291
status.operation.instrument.trigger_timer.trigger_overrun.*
This attribute contains the operation status trigger timer overrun register set.
Type TSP-Link accessible Affected by Where saved Default value
Usage
operationRegister =
status.operation.instrument.trigger_timer.trigger_overrun.condition
operationRegister =
status.operation.instrument.trigger_timer.trigger_overrun.enable
operationRegister =
status.operation.instrument.trigger_timer.trigger_overrun.event
operationRegister =
status.operation.instrument.trigger_timer.trigger_overrun.ntr
operationRegister =
status.operation.instrument.trigger_timer.trigger_overrun.ptr
status.operation.instrument.trigger_timer.trigger_overrun.enable =
operationRegister
status.operation.instrument.trigger_timer.trigger_overrun.ntr =
operationRegister
status.operation.instrument.trigger_timer.trigger_overrun.ptr =
operationRegister
The status of the operation status trigger timer trigger overrun register; a zero (0)
indicates no bits set (also send 0 to clear all bits); other values indicate various bit
Details
These attributes are used to read or write to the operation status trigger timer overrun registers. Reading a status
register returns a value. The binary equivalent of the returned value indicates which register bits are set. The
least significant bit of the binary number is bit B0, and the most significant bit is bit B15. For example, if a value
of 18 is read as the value of the condition register, the binary equivalent is 0000 0000 0001 0010. This value
indicates that bit B1 and bit B4 are set.
* Least significant bit
** Most significant bit