Section
7: TSP command reference Series 2600B System SourceMeter® Instrument
Reference Manual
7-230 2600BS-901-01 Rev. B / May 2013
smuX.source.rangeY
This attribute contains the source range.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW)
Yes
SMU reset
Instrument reset
Recall setup
Saved setup
Voltage:
Models 2601B/2602B/2604B: 100e-3 (100 mV)
Models 2611B/2612B/2614B/2634B/2635B/2636B:
200e-3 (200 mV)
Current:
Models 2601B/2602B/2604B/2611B/2612B/2614B:
100e-9 (100 nA)
Models 2634B/2635B/2636B: 1e-9 (1 nA)
Usage
rangeValue = smuX.source.rangeY
smuX.source.rangeY = rangeValue
rangeValue
Set to the maximum expected voltage or current to be sourced
X
Source-measure unit (SMU) channel (for example, smua.measure.rangev
applies to SMU channel A)
Y
SMU source function (v = voltage, i = current)
Details
This attribute contains a value that sets the source-measure unit (SMU) to a fixed range large enough to source
the value. When read, the attribute contains the range the instrument is presently on when in autorange.
Assigning a value to this attribute sets the SMU to a fixed range large enough to source the assigned value. The
instrument selects the best range for sourcing a value of rangeValue.
Reading this attribute returns the positive full-scale value of the source range the SMU is currently using. With
source autoranging enabled, the output level controls the range. Querying the range after the level is set returns
the range the instrument chose as appropriate for that source level.
This attribute is primarily intended to eliminate the time required by the automatic range selection performed by a
sourcing instrument. Because selecting a fixed range prevents autoranging, an overrange condition can occur.
For example, sourcing 10.0 V on the Model 2601B/2602B/2604B 6 V range or sourcing 5.0 V on the Model
2611B/2612B/2614B 2 V range causes an overrange condition.
Example
smua.source.rangev = 1
Selects the 1 V source range for SMU channel A.
Also see
Range (on page 2-82)
reset() (on page 7-160)
setup.recall() (on page 7-180)
smuX.measure.rangeY (on page 7-210)
smuX.reset() (on page 7-216)
smuX.source.autorangeY (on page 7-217)

Table of Contents

Other manuals for Keithley 2601B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2601B and is the answer not in the manual?

Keithley 2601B Specifications

General IconGeneral
BrandKeithley
Model2601B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals