7: TSP command reference Series 2600B System SourceMeter® Instrument
7-156 2600BS-901-01 Rev. B / May 2013
printnumber()
This function prints numbers using the configured format.
Type TSP-Link accessible
Affected by Where saved Default value
Usage
printnumber(value1)
printnumber(value1, value2)
printnumber(value1, ..., valueN)
First value to print in the configured format
Second value to print in the configured format
Last value to print in the configured format
One or more values separated with commas
Details
There are multiple ways to use this function, depending on how many numbers are to be printed.
This function prints the given numbers using the data format specified by format.data and
format.asciiprecision.
Example
format.asciiprecision = 10
x = 2.54
printnumber(x)
format.asciiprecision = 3
printnumber(x, 2.54321, 3.1)
Configure the ASCII precision to 10 and set
to
2.54.
Read the value of x based on these settings.
Change the ASCII precision to 3.
View how the change affects the output of x and
some numbers.
Output:
2.540000000e+00
2.54e+00, 2.54e+00, 3.10e+00
Also see
format.asciiprecision (on page 7-92)
format.byteorder (on page 7-93)
format.data (on page 7-94)
print() (on page 7-152)
printbuffer() (on page 7-153)
PulseIMeasureV()
This KIPulse factory script function performs a specified number of pulse I, measure V cycles.
Type TSP-Link accessible
Affected by Where saved Default value