System SourceMeter® Instrument Reference Manual Section 7:
2600BS-901-01 Rev. B / May 2013 7-51
digio.trigger[N].assert()
This function asserts a trigger on one of the digital I/O lines. This function is not available on the Models
2604B/2614B/2634B.
Type TSP-Link accessible
Affected by Where saved Default value
Usage
digio.trigger[N].assert()
Digital I/O trigger line (1 to 14)
Details
The set pulsewidth determines how long the trigger is asserted.
Example
digio.trigger[2].assert()
Asserts a trigger on digital I/O line 2.
Also see
digio.trigger[N].pulsewidth (on page 7-54)
digio.trigger[N].clear()
This function clears the trigger event on a digital I/O line. This function is not available on the Models
2604B/2614B/2634B.
Type TSP-Link accessible
Affected by Where saved Default value
Usage
digio.trigger[N].clear()
Digital I/O trigger line (1 to 14)
Details
The event detector of a trigger enters the detected state when an event is detected. It is cleared when
digio.trigger[N].wait() or digio.trigger[N].clear() is called.
digio.trigger[N].clear() clears the event detector of the specified trigger line, discards the history of the
trigger line, and clears the digio.trigger[N].overrun attribute.
Example
Clears the trigger event detector on I/O line 2.
Also see
digio.trigger[N].overrun (on page 7-54)
digio.trigger[N].wait() (on page 7-58)