Series 2600B
System SourceMeter® Instrument Reference Manual Section 7:
TSP command reference
2600BS-901-01 Rev. B / May 2013 7-55
Example
digio.trigger[4].pulsewidth = 20e-6
Sets the pulse width for trigger line 4 to
20 μs.
Also see
digio.trigger[N].assert() (on page 7-51)
digio.trigger[N].reset() (on page 7-55)
digio.trigger[N].release() (on page 7-55)
digio.trigger[N].release()
This function releases an indefinite length or latched trigger. This function is not available on the Models
2604B/2614B/2634B.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
digio.trigger[N].release()
N
Digital I/O trigger line (1 to 14)
Details
Releases a trigger that was asserted with an indefinite pulse width time. It also releases a trigger that was
latched in response to receiving a synchronous mode trigger. Only the specified trigger line is affected.
Example
digio.trigger[4].release()
Releases digital I/O trigger line 4.
Also see
digio.trigger[N].assert() (on page 7-51)
digio.trigger[N].pulsewidth (on page 7-54)
digio.trigger[N].reset()
This function resets trigger values to their factory defaults. This function is not available on the Models
2604B/2614B/2634B.
Type TSP-Link accessible
Affected by Where saved Default value
Function Yes
Usage
digio.trigger[N].reset()
N
Digital I/O trigger line (1 to 14)

Table of Contents

Other manuals for Keithley 2601B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2601B and is the answer not in the manual?

Keithley 2601B Specifications

General IconGeneral
BrandKeithley
Model2601B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals