7: TSP command reference Series 2600B System SourceMeter® Instrument
7-180 2600BS-901-01 Rev. B / May 2013
setup.recall()
This function recalls settings from a saved setup.
Type TSP-Link accessible
Affected by Where saved Default value
Usage
setup.recall(id)
An integer or string that specifies the location of the setup to recall:
• Factory default setup (0)
• User-saved setup in nonvolatile memory (1 to 5)
• User-saved setup on a USB flash drive (
"/path/filename"
Details
When the id parameter is an integer (n), it is interpreted as the setup number to restore from the instrument's
nonvolatile memory. When n = 0, the instrument recalls the factory default setup; when n = 1 to 5, the
instrument recalls a user-saved setup.
When the id parameter is a string, it is interpreted as the path and file name of the setup to restore from a file on
a USB flash drive. The path may be absolute or relative to the current working directory.
Before a setup is recalled, an instrument reset is performed.
Example 1
Recall the user-saved setup at location 1.
Example 2
setup.recall("/usb1/KEITHLEY_30730.set")
Recall a user-saved setup stored in a file
named KEITHLEY_30730 on a USB flash
Also see
setup.save() (on page 7-181)
Saved setups (on page 2-46)