System SourceMeter® Instrument Reference Manual Section 7:
TSP command reference
2600BS-901-01 Rev. B / May 2013 7-319
As an example, to set bit B0 of the standard event status enable register, set status.standard.enable =
status.standard.OPC.
In addition to the above constants, standardRegister can be set to the numeric equivalent of the bit to set. To
set more than one bit of the register, set standardRegister to the sum of their decimal weights. For example,
to set bits B0 and B4, set standardRegister to 17 (which is the sum of 1 + 16).
Bit
B7
B6
B5
B4
B3
B2
B1
B0
Binary value
0/1
0/1
0/1
0/1
0/1
0/1
0/1
0/1
Decimal
128
64
32
16
8
4
2
1
Weights
(2
7
) (2
6
) (2
5
) (2
4
) (2
3
) (2
2
) (2
1
) (2
0
)
Example 1
standardRegister = status.standard.OPC
+ status.standard.EXE
status.standard.enable = standardRegister
Uses constants to set the OPC and EXE
bits of the standard event status enable
register.
Example 2
-- decimal 17 = binary 0001 0001
standardRegister = 17
status.standard.enable = standardRegister
Uses a decimal value to set the OPC and
EXE bits of the standard event status
enable register.
Also see
Standard Event Register (on page E-20)
status.system.*
These attributes manage the TSP-Link
®
system summary register of the status model for nodes 1 through 14.
These attributes are not available on the Models 2604B/2614B/2634B.
Type TSP-Link accessible Affected by Where saved Default value
Attribute
- -
- -
- -
- -
.condition (R) Yes Not applicable Not saved Not applicable
.enable (RW)
Yes
Status reset
Not saved
0
.event (R)
Yes
Status reset
Not saved
0
.ntr (RW) Yes Status reset Not saved 0
.ptr (RW)
Yes
Status reset
Not saved
32,767 (All bits set)
Usage
enableRegister = status.system.condition
enableRegister = status.system.enable
enableRegister = status.system.event
enableRegister = status.system.ntr
enableRegister = status.system.ptr
status.system.enable = enableRegister
status.system.ntr = enableRegister
status.system.ptr = enableRegister
enableRegister
The status of the system summary register; a zero (0) indicates no bits set; other
values indicate various bit settings

Table of Contents

Other manuals for Keithley 2601B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2601B and is the answer not in the manual?

Keithley 2601B Specifications

General IconGeneral
BrandKeithley
Model2601B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals